Operating Mode: Transmittance (%), Absorbance, and Concentration
Source Lamp: Tungsten (visible) and Deuterium (UV)
Wavelength Range: 190 - 1100 nm
Wavelength Accuracy: ¡¾ 1 nm in Wavelength Range 200 - 900 nm
Wavelength Resolution: Selectable (0.1, 0.2, 0.5, 1.0, 2.0, and 5.0 nm)
Wavelength Calibration: Automatic
Wavelength Selection: Automatic, based on method selection
Scanning speed: Depending on selected resolution (100-4500 nm/min.)
Spectral Bandwidth: 3 nm
Photometric Readout: -0.3 to 3.0 A or 0.1 to 100%T
Photometric Accuracy: ¡¾ 0.005 A at 0.0 to 0.5 A 1% at 0.5 to 2.0 A
Photometric Linearity: <0.5% at 2.0 A 1% at > 2.0 A
Stray Light: KI Solution at 220 nm > 3.3 Abs
DU 700 is < o r = 0.05 %T max at 220 and 320 nm
Size: W450 x H200 x D500mm
Weight: 15.5 kg
Operating Requirements: 10 - 40¡ÆC, max 90% relative humidity (non-condensing)
Storage Requirements: -25 - 60¡ÆC, max 80% relative humidity (non-condensing)
Power: 100 - 240 VAC, 50/60 Hz, automatic changeover
Ports: USB 1.1
Water-resistant to : IP 32